Carrier Calibration Wizard
The Carrier Calibration Wizard gives you a simple way to adjust the calibration on a Carrier input, without needing to go into Configuration — a section you generally shouldn't need to enter for routine calibration adjustments.
Step-by-Step Procedure
Section titled “Step-by-Step Procedure”Navigate to Wizards: From Manual Mode, switch to the Wizards tab.
Select the Wizard: Choose Calibrate Carrier Flow Meter from the Wizard to Run dropdown.
Select the Input: If you have more than one Carrier input, select the correct one.
Enter Amount Mixmate Output: Enter the amount the machine reports it dispensed.
Enter Verified Actual Amount Output: Enter the amount you physically measured coming out.
Compare: Review the previous calibration against the newly calculated calibration.
Save: Press Save to apply the calibration change.