Troubleshooting
Flowmeter Calibration
The carrier meters that are used with the system typically have to be calibrated after final installation because they are sensitve to how the liquid flows through them. We typically like to see how the total comes out over a period of 3 loads before ...
Calibration of Inductors
How to verify inductor calibration: 1. Go to manual mode (Go to Data > Mixmates > Tap on icon next to the desired Mixmate) 2. Turn on scale reading for inductor 3. Tare the weight so it says ~0 4. Measure in 2.5 gallons of water 5. Write down what ...
Tablet Error Codes
System Level Errors 101 - Unexpected Flow - Detects flow through a non used measurement device 102 - Failed SRAM - Master Controller memory chip failure, need new module 103 - Emergency Stop Request from Machine - Something caused emergency stop to ...
Power Module Blink Code Diagnostics
The power module now has blink codes. The number of flashes happens once every 5 seconds. The power must be cycled to clear the code. 1 blink - Low Voltage Dropout ( resets below approximately 10.5 volts ) 2 blinks - Master Controller unexpected ...
Terminators Required
Two (2) blue* Terminators are REQUIRED to be connected to your Mixmate CANBUS for your system to run properly. They should be placed as far apart from each other as possible. (opposite ends of the CANBUS) Pro Inductors systems have one installed ...
Jug Rinse Sensor Replacement
Load Cell Connector Replacement
Slow mix time with thicker chemical(s)
When using a highly viscous chemical, it is possible to add a pump at the tote, so it can push the product into the Mixmate and move things along quicker. Make sure that the overall flowrate stays below 35 gal/min as to not damage gear meter. Lastly, ...