ARAG Orion X 3" Mag Meter Manual

ARAG Orion X 3" Mag Meter Manual

ARAG Orion X 3" Magmeter Specifications & Installation

The Orion X is an electromagnetic flowmeter designed for agricultural applications
. Because it is electromagnetic, it contains no moving parts, making it ideal for chemicals, provided they meet the minimum conductivity requirements.

1. Technical & Metric Performance

  • Measuring Range: 50 – 2500 L/min (approx. 13.2 – 660.4 GPM)

  • Accuracy: +/- 1% of the measured value (within the 250 to 2500 L/min range)

  • Maximum Pressure: 5 bar (approx. 72.5 PSI)

  • Liquid Conductivity: Must be equal to or greater than 300 µS/cm

  • Operating Temperature: 0°C to +60°C (32°F to +140°F)

2. Hydraulic Installation Logic

To ensure measurement accuracy and prevent turbulence, follow these placement rules:

  • The 10x Rule: Install a straight pipe section both upstream and downstream of the meter. The length must be at least 10 times the inner diameter

    • For this 3" (75mm) meter, that is 750 mm (approx. 29.5 inches) of straight pipe on both sides

  • Avoid Turbulence: Position the meter away from valves, bends, or bottlenecks

  • Air Pocket Prevention: Always mount the meter in a position that favors the drainage of air

  • Electrode Protection: Install the meter at least 3 cm (approx. 1.2 inches) away from metal masses

3. Electrical Connections & Pinouts

The Orion X features two ports: Frequency and CAN Bus. Use the following pinout for wiring to your Mixmate modules:

Frequency Port (Pulse Output):

  • Pin 1: GND (Ground)

  • Pin 2: +12 Vdc (Power)

  • Pin 3: OUT (Square wave pulse signal)

4. Status LED Indicators

The multi-color LED on the back of the meter provides instant diagnostic feedback:

  • Blue (Steady): Measurement active; liquid is flowing
  • Green (Steady): Liquid is present but still (no flow)
  • Violet (Steady): No liquid detected inside the meter
  • White (Steady): Electrode cleaning in progress
  • Red (Blinking): Error detected (refer to the pulse count for specific error codes)

5. Maintenance

  • Post-Mix Flush: Always run clean water through the meter after each treatment session

  • Electrode Cleaning: If the reading becomes unstable, use a special detergent to clean the pipe. Do not use metal or abrasive objects as they will damage the internal electrodes


Attached is the english version of the manual. 

Here is the link to the other versions of the manual https://www.aragnet.com/EN/USA/246/products/?resetData=1&PRD=122013